7
(2) Unload the moving board 1 when observe culture tissue in large dish. When unload, please put condenser system 4
out of the light path, take out the dish board and lower the objective. Unscrew the screw 5, place upside down on
working platform, it can not put sidelong, or it might be damaged . See figure 10.
(3) Tightness adjustment
It has been adjusted well from factory. If needed after long time using, unload the ruler, reverse it and do the steps
below.
(a) Y direction
Unscrew screw 4, unload the supporting board 3, unscrew the two screws 1 (a little bit) and then adjust screw 2 with
Allen wrench to adjust the tightness of Y transmission steel rope. See figure 11-a.
(b) X direction
Unscrew the two screws 5 and adjust screw 6 to adjust the tightness of X transmission steel rope. See figure 11-b.
Moving stage ruler is the key parts to hold slider and sample, it can not put
sidelong or without anything to support it on back after unload, please put it upside
down stable, or it be damaged on accuracy.
Notice
Figure 11
4
3
6
2
a
b
5
1
Figure
10
5
4