background image

Check Status of Logger:

During Recording [RUN] & Stop [STOP] mode, press START button once to view MIN | MAX | AVERAGE values

Display Indication:

Only Current Reading 

Not Started

START     

Waiting to START

RUN  

Started Recording 

STOP    

Recording is Stop

OK     

No Alarm Condition 

NOT OK     

LO  | HI Alarm Condition

Stop Logger:

∙ Keep STOP button pressed for 5 seconds to stop logger
∙ Display will indicate STOP confirming logger has been stopped successfully 

Note: Logger will be stop after 10 seconds when connected with USB port of PC

PDF Generation:

∙ After stopping, when the logger is connected for first time to USB port of the computer, both LEDs (GREEN & 

RED) will start flashing to show the PDF generation is in process
∙ 

Do Not Disconnect Logger until both LEDs (GREEN & RED) are ON continuously

, which indicates PDF is ready

∙ Then folder containing PDF will open automatically. Double click to open Copy and Save As PDF file to desired 
location 
∙ Once the PDF is generated, whenever the logger is connected to USB port of the computer, both LEDs (GREEN & 
RED) are ON continuously, which indicates logger is in stop mode and PDF is already generated

sales@msr.ch

www.msr.ch

BudgetLine

PDF USB Loggers

Quick Start Guide