25
POR
TUGUÊS
S.39.17
FOREIGN MATERIAL
When the
FOREIGN MATERIAL
function is
enabled, the equipment will ask if the user
wishes to perform this function every time at the
beginning of the measurement cycle.
• IMPORTANT
: In the process of measuring
the foreign material,
keep away the dump
cell
. The sample should be added directly
into the grain cell.
• IMPORTANT
: The reference sample shall
have a minimum weight of
250,0 grams
.
• Set the empty grain cell on the moisture
meter.
• Place the full reference sample in the grain
cell. Wait for stabilization.
• When prompted, remove the sample, return
the empty grain cell into the equipment and
press
OK
.
• Separate all foreign material from the full
sample.
• Place in the grain cell everything that has
been separated from the full sample.
• The moisture meter will show the
percentage of foreign material in relation to
the full sample.
• Press
Continue
.
After measuring the foreign material, the
moisture reading as described on
page 19
.
FOREIGN MATERIAL AND DAMAGED KERNEL