HYDROCAL 1005 Manual for Installation and Operation
Page 99/111
5.13.3
Alarm window
The alarm window shows all alarm status activated for presentation. The results are presented in
form of a time trend chart or a table. When on the alarm window, click right mouse button to open
the menu and select the form desired. There is also the possibility to activate the print-function (only
from time trend chart) or export function (only from table view) and to fade out the sensor and time
range selection.
5.13.4
Status of alarms
Five different alarm status are presented in both “View as table” and in “View as graph”. The mean-
ing of the different status is described as follows:
Status 0:
No alarm active, the limits for the alarm have not be exceeded in the past or an
alarm existing has been acknowledged.
Status 1:
The limit defined for a High - Alarm was exceeded in the past, but value has fallen
below the limit.
Status 2:
The defined “Level 2“ for a High - High - alarm has been exceeded in past. The value
has fallen below the Level 2, but is still higher than „Level 1“.
Status 3:
The limit defined for a High - Alarm is currently exceeded.
Status 4:
The defined level for a High - High - Alarm is currently exceeded.
5.13.5
Print and export of results
There exist two possibilities to process measuring data for the purpose of documentation:
From the table view, data can be exported to MS Excel © or as a text file. Export to Excel can be
activated either
from the menu or by „File->Export->Excel…“. Export into a text file is possible
only by „File->Export->CSV…“.