HYDROCAL 1005 Manual for Installation and Operation
Page 51/111
Out, hold & SMS:
An entry in the history of alarms is made; this has to be confirmed (see history
of alarms) and an output is switched on (If the measured value exceeded the configured limits in
the field "Level 1 & 2")
. The status „on“ described above is activated. If the measured value for
this output falls under the limit, the output remains switched on. Only a manual deleting of alarm
allows that the output is switched off again (see chapter 4.11 or 3.6). In addition, an SMS is sent
to the number indicated.
If it is possible to use the same output for different alarms and therefore, collection alarms can be
defined.
4.9.5
Unit Setup
The following setup can be modified:
Digital Output Test
Device Options setup (Backlight display)
Date. Time & Clock Settings
Communication settings
Exit