background image

                                                                                               

 

1

 

ETHOS 5030 

TRUE-RMS DIGITAL MULTIMETER 

OPERATION MANUAL 

 

 
 
 

Summary of Contents for ethos 5030

Page 1: ...1 ETHOS 5030 TRUE RMS DIGITAL MULTIMETER OPERATION MANUAL ...

Page 2: ...he following safety information must be observed to insure maximum personal safety during the operation at this meter 1 1 Do not operate the meter if the body of meter or the test leads appear damaged 1 2 Check the main function dial and make sure it is at the correct position before each measurement 1 3 When making current measurements ensure that the circuit not live before opening it in order t...

Page 3: ...andards EMC LVD CAT Ⅲ 600V CAT Ⅳ 300V The meter is up to the standards of IEC1010 Double Insulation Pollution Degree 2 Overvoltage Category Ⅲ Operating environment Temperature 32 to104 F 0 C to 40 C Humidity 80 RH Storage environment Temperature 4 to140 F 20 C to 60 C Humidity 90 RH Fuse F0 8A 600V 5 x 20mm F15A 600V 5 x 20mm Power supply 9V battery 6F22 Test lead probe RE10 CATⅢ 600V CATⅣ 300V 15...

Page 4: ... 600mA 100µA 6A 2 0 3 1mA 15A 10mA Overload protection 0 8A 600V 15A 600V fuse Continuous testing the maximum steady state current of 1A 2 2 4 AC Current Range Accuracy Resolution 600µA 1 5 3 0 1µA 6000µA 1µA 60mA 10µA 600mA 100µA 6A 2 5 5 1mA 15A 10mA AC True RMS Frequency 40Hz 1kHz Overload protection 0 8A 600V 15A 600V fuse Continuous testing the maximum steady state current of 1A 2 2 5 Resista...

Page 5: ...2V Built in buzzer sounds if resistance is less than 50Ω Open circuit voltage approx 1V Overload protection 600V DC or AC rms 2 2 8 Frequency Range Accuracy Resolution 9 999Hz 0 1 5 0 001Hz 99 99Hz 0 01Hz 999 9Hz 0 1Hz 9 999kHz 1Hz 99 99kHz 10Hz 999 9kHz 100Hz 9 999MHz 1kHz Sensitivity sine wave 0 6V rms 9 999MHz 1 5V rms Overload protection 600V DC or AC rms 2 2 9 Duty cycle 0 1 99 9 2 0 2 Freque...

Page 6: ...easurement is to be taken Connect the meter in series with the circuit 5 Read the result from the LCD panel 6 On AC range press Hz key to measurement frequency or duty cycle 3 3 Resistance Measurement 1 Connect the black test lead to COM socket and red test lead to the VΩHz socket 2 Set the selector switch to desired position 3 Connect tip of the test leads to the points where the value of the res...

Page 7: ...3 7 Temperature Measurement 1 Connect the black banana plug of the sensor to COM socket and red banana plug to the VΩHz socket 2 Set the selector switch to desired position 3 Press SELECT key to choose or measurement 4 Put the sensor probe into the temperature field under measurement 5 Read the result from the LCD panel 3 8 Data Hold On any range press the key to lock display value and the DH sign...

Page 8: ...replacement 1 This meter is provided with a 0 8A 600V fuse to protect the battery test and the current measuring circuits which measure up to 600mA with a 15A 600V fuse to protect the 15A range 2 Ensure the instrument is not connected to any external circuit set the selector switch to OFF position and remove the test leads from the terminals 3 Remove the three screws on the bottom case and lift th...

Page 9: ...9 Distributed by MTi Victoria Road Avonmouth Bristol BS11 9DB Tel 0117 938 6400 Fax 0117 923 5374 Email enquiries ethos adivision co uk Web www ethos instruments co uk ...

Reviews: