General Specifications
- Power Supply: 21.5V alkaline batteries (LR6)
- Battery Life: about 400mA/3V under the condition
of 10mA with 1kΩ load
- Max. permitted Voltage: 30V(between any two
terminals or between any terminal and earth
ground)
- Operating temperature: 0
℃
to 50
℃
- Operating relative humidity: ≤ 80% RH
- Storage temperature: ≤ -10
℃
to 50
℃
- Relative humidity for storage: ≤ 90% RH
- Temperature co-efficiency: 0.1×
(
designated
accuracy
)
%/
℃(
5
℃~
18
℃、
28
℃~
40
℃)
- Size: 180
(
L
)
×90
(
W
)
×47
(
D
)
mm (with
holster)
- Weight: 500g(with holster)
- Accessories: Operation instruction, a set of CF-36
industrial test lead(with alligator clips attached to
probes)
- Safety: Certified as compliant to IEC1010
provisions(Safety standard issued by International
Electrotechnical Commission)
Section Nine: Calibration
Note
Calibration: In order to keep the designed
accuracy of this Instrument, it is recommendable to
calibrate your calibrator once a year. The following
example shows the use of recommended standard
equipment to perform the calibration.
Caution
Usage
:
Never apply more than the max.
permitted voltage to the input of the calibrator,
otherwise the overvoltage will lead to possible
damage to the input section.
Usage: during the operation, avoid short
circuit and never apply more than the max.
permitted voltage to the output of the Instrument or
to a co-working standard device, otherwise any
mal-operation will cause possible damage to their
internal circuits.