Selecting Standard Equipment
Table 10-1
Calibration Item
Standard
Equipment
Output Range
Accuracy
Recommended
Output
DCA 20mA
Digit meter
MAX.22 mA
±
(
50ppm
+
0.4uA
)
Digit meter:KEITHLEY 2000
Standard resistance:BZ10-100Ω
Measure
ment
DCA
20mA
Standard
source
MAX.33mA
±
(
100ppm
+
0.2uA
)
5520 (FLUKE) or equivalent
DCV 28V
Standard
source
MAX.33V
±
(
12ppm
+
15uV
)
28
29