CAUTION
Improper use of this meter can cause damage, shock, injury or death. Read and
understand this user’s manual before operating the meter.
Always remove the test leads before replacing the battery.
Inspect the condition of the test leads and the meter itself for any damage before operating
the meter. Repair or replace any damage before use.
Use great care when making measurements if the voltages are greater than 25VAC rms or
35VDC. These voltages are considered a shock hazard.
Remove the battery if the meter is to be stored for long periods.
Always discharge capacitors and remove power from the device under test before
performing Diode, Resistance or Continuity tests.
Voltage checks on electrical outlets can be difficult and misleading because of the
uncertainty of connection to the recessed electrical contacts. Other means should be used
to ensure that the terminals are not live.
If the equipment is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
Input limits
Function
Maximum input
A AC
400A
V DC, V AC
600V DC/AC
Frequency, Resistance, Diode,
Continuity, Capacitance
600V DC/AC
Temperature (ºC/ºF)
600V DC/AC
2