MControl Can’t Connect to the Internet When Using an EDR 

Workstation 

1.

 

First check to see if you are connecting through an EDR’s Internet service.  Most 

companies will provide this through their workstation.

  

The cable that the

 

EDR’s 

workstation uses should be a CAT 5 Ethernet connection and will connect to the back of 

the workstation.  There will most often be a second Ethernet port in the back of the 

workstation.  Take another Ethernet cable and connect it to the empty port on the 

workstation.  Connect the other end to the MLogger at the red WAN port located at the 

back of MLogger.  Make sure your computer is connected to MLogger through a blue 

port for a local connection.  Allow some time to pass so an address can be located with 

the EDR’s server.  Next open Internet Explorer and confirm that you now have internet.

 

2.

 

When using a Pason workstation it is possible to use the MLogger’s router like a switch.  

Plug an Ethernet cable into the Workstation’s second network port and the other end into 

a blue port in MLogger.  Connect your computer’s Ethernet cable into another one of 

MLogger’s blue ports.  Open Internet Explorer and access a web page to confirm.

 

Filament Baseline is drifting on MControl after zeroing them 

1.

 

This is caused by a temperature change in the atmosphere.  The MLogger has a 

heated detector block that houses the filaments.  This block is kept at a steady 120 

degrees Fahrenheit to help keep the sample gas at a constant temperature.  However, 

even with the heated detector block the base line will have a noticeable drift from 

daylight to nighttime because of the natural temperature change.

 

2.

 

Check the PID controller for activity.  It’s located on the 

Blue side

.  The display should 

read 120 degrees Fahrenheit, or close to it.  If there is no activity, switch to the spare 

MLogger and contact the MSI office. 

100% Methane Test Gas Calibration Does Not Reach 10,000 Units 

1.

 

Check that all exhaust lines are connected.  The main exhaust must be connected to 

provide enough backpressure for a good sample.  The total gas units may be off by 

2000 units if there is no main exhaust line connected to the correct poly-flow fitting.

 

2.

 

 Check the TCD and CCD filaments; there may be a leak where it seals to the detector 

block.  Make sure the screws are snug; do not over tighten the screws.  There should 

Summary of Contents for MLogger Blue Box

Page 1: ...he 1 Test Gas 14 Filaments won t Adjust from a Negative Number to a Positive Number 15 Changing a Filament 15 MLogger Main Flow won t Adjust Higher than 2 SCFH 15 MLogger Carrier Air Pressure won t Ad...

Page 2: ...aff at MSI may need you to locate certain parts on the unit to help troubleshoot the problem The MLogger is a complex and sophisticated machine Before attempting to fix any internal problems please fa...

Page 3: ...3 This is the right side of the box when facing front of MLogger Blue Side...

Page 4: ...4 This is the front view of the MLogger Model TGC This is the back view of MLogger...

Page 5: ...5 Sample Flow Diagram...

Page 6: ...while the other will flash If there are no lights on the router make sure all the Ethernet cables are plugged in securely a When connected to one of the blue ports try unplugging and reconnecting whi...

Page 7: ...nd left click on the properties button g Check to see if the Obtain an IP Address Automatically has a bullet mark next to it If it doesn t then this needs to be done Also be sure that the Obtain DNS S...

Page 8: ...on an office network using a server Some networks have strong firewall and antivirus programs that may not be disabled Check with the IT department in the office and ask them to allow ports 20 throug...

Page 9: ...rocedure 6 Reset the thresholds to the original values 480 for CCD threshold and 440 for TCD threshold 7 Turn the ball valve on the blue side from Zero back to Rig If the problem persists then you may...

Page 10: ...tting the router to obtain addresses automatically DHCP can fix the problem right away and is the best thing to try first LINKSYS WIRELESS ROUTER 1 open Internet Explorer and type 192 168 123 254 in t...

Page 11: ...lite cable directly to your computer The router in the MLogger needs to be reconfigured to recognize the satellite s modem A First find the IP address for the satellite i Take the satellite Ethernet c...

Page 12: ...window Now this next orange window will have a heading of WAN IP Address Assignment This is where you need to enter the new IP address IP Subnet Mask and Gateway IP Address obtained from step Ai Aiii...

Page 13: ...ernet cable into another one of MLogger s blue ports Open Internet Explorer and access a web page to confirm Filament Baseline is drifting on MControl after zeroing them 1 This is caused by a temperat...

Page 14: ...ed Side and inspect the IT pump Be sure all lines are connected If all the lines are connected then there is a leak in the pump itself This requires the pump to be replaced Switch to the spare MLogger...

Page 15: ...the connector and pull There are no clasps or latches holding the plug in place 3 Use a small Phillips screwdriver to remove the screws holding the filament in place Remove the old filament 4 Insert t...

Page 16: ...op of the flow meter Use a pipe cleaner to clean and free up the bead c Or try this i Disconnect all the power serial and exhaust lines from MLogger Once the box is free from any tether turn the unit...

Page 17: ...low diagram b If all fittings are tight and no leaks are found then the regulator needs to be replaced Switch to the spare MLogger and contact the MSI office 3 Look inside the MLogger on the Red Side...

Page 18: ...is carrier air pressure and the GC flow is at 0 5 SCFH then the Chromatograph Valve Manifold may be malfunctioning Check the 1 8 poly flow fittings for any loose connections or holes in the line If no...

Page 19: ...ence from an electromagnetic field a Check for any fuzz buster HAM radio antenna or satellite phone in operation close by b Cell phones in use next to the MLogger can sometimes affect the filament sig...

Page 20: ...toolbar ii Check the WITS Input bullet is marked iii Check the WITS Interface Be sure the correct EDR company shows in the white field iv Check the WITS port for the correct connection COM 1 in the fr...

Page 21: ...the sever should cache this connection select yes d The black window will then display the various WITS records that MLogger is receiving from the EDR e The TAG column in this screen corresponds to t...

Page 22: ...lashing green lights on both the processor and router then everything may be working fine The small LED for the hard drive light may be broken or has become unplugged from the processor This will not...

Page 23: ...d drive 1 Check how full the Log file and the Chromatograph file are a Look to the Status window in MControl If the LF and the CF are both at 100 the data has wrapped and MLogger is over writing data...

Reviews: