background image

 

17 

 

b.

 

Remove the Filter Housing by loosening both poly flow fittings with a 7/16” 

wrench. 

c.

 

Open the Filter Housing using two 9/16” wrenches on opposite ends.  When the 

housing comes loose be careful not to lose the small spring located within (see 

picture below). 

 

 

 

 

d.

 

Replace the filter and make sure the number “15” is showing face up when placed 

inside the housing.  Followed by the spring and then the top will screw back into 

place. 

e.

 

Tighten the housing with the 9/16” wrenches and replace the Filter Housing back 

in line with the Carrier Air port using the 7/16” poly low fittings. 

2.

 

If the Carrier Air still won’t adjust after changing the 15 Micron filter, then the 

Regulator may be bad.   

a.   Check all fittings by tracing the flow using the 

flow diagram

 

b.

 

If all fittings are tight and no leaks are found, then the regulator needs to be 

replaced.  Switch to the spare MLogger and contact the MSI office.  

3

 

Look inside the MLogger on the 

Red Side

.  Identify the 24-volt pump and feel if there is 

any vibration in the pump.  If the pump is quiet and still, then it has gone bad and 

needs replaced.  Switch to the spare MLogger and contact the MSI office.  

Chromatograph won’t Display any Gas Peaks 

1.

 

Re-Zero the GC filament and try another chromatograph run.  

Note: do not re-zero 

the GC when the chromatograph is running; be sure the chromatograph is 

stopped

.  

2.

 

Check the scale used to measure the chromatograph window.   Verify that the correct 

scale has been selected.  Small curves require a small scale to observe.   

3.

 

Check the GC flow meter to see if it will adjust.  If it doesn’t adjust, the ball valve may 

be turned to Test Gas.  Or, there may be a leak before the sample reaches the flow 

meter. 

Summary of Contents for MLogger Blue Box

Page 1: ...he 1 Test Gas 14 Filaments won t Adjust from a Negative Number to a Positive Number 15 Changing a Filament 15 MLogger Main Flow won t Adjust Higher than 2 SCFH 15 MLogger Carrier Air Pressure won t Ad...

Page 2: ...aff at MSI may need you to locate certain parts on the unit to help troubleshoot the problem The MLogger is a complex and sophisticated machine Before attempting to fix any internal problems please fa...

Page 3: ...3 This is the right side of the box when facing front of MLogger Blue Side...

Page 4: ...4 This is the front view of the MLogger Model TGC This is the back view of MLogger...

Page 5: ...5 Sample Flow Diagram...

Page 6: ...while the other will flash If there are no lights on the router make sure all the Ethernet cables are plugged in securely a When connected to one of the blue ports try unplugging and reconnecting whi...

Page 7: ...nd left click on the properties button g Check to see if the Obtain an IP Address Automatically has a bullet mark next to it If it doesn t then this needs to be done Also be sure that the Obtain DNS S...

Page 8: ...on an office network using a server Some networks have strong firewall and antivirus programs that may not be disabled Check with the IT department in the office and ask them to allow ports 20 throug...

Page 9: ...rocedure 6 Reset the thresholds to the original values 480 for CCD threshold and 440 for TCD threshold 7 Turn the ball valve on the blue side from Zero back to Rig If the problem persists then you may...

Page 10: ...tting the router to obtain addresses automatically DHCP can fix the problem right away and is the best thing to try first LINKSYS WIRELESS ROUTER 1 open Internet Explorer and type 192 168 123 254 in t...

Page 11: ...lite cable directly to your computer The router in the MLogger needs to be reconfigured to recognize the satellite s modem A First find the IP address for the satellite i Take the satellite Ethernet c...

Page 12: ...window Now this next orange window will have a heading of WAN IP Address Assignment This is where you need to enter the new IP address IP Subnet Mask and Gateway IP Address obtained from step Ai Aiii...

Page 13: ...ernet cable into another one of MLogger s blue ports Open Internet Explorer and access a web page to confirm Filament Baseline is drifting on MControl after zeroing them 1 This is caused by a temperat...

Page 14: ...ed Side and inspect the IT pump Be sure all lines are connected If all the lines are connected then there is a leak in the pump itself This requires the pump to be replaced Switch to the spare MLogger...

Page 15: ...the connector and pull There are no clasps or latches holding the plug in place 3 Use a small Phillips screwdriver to remove the screws holding the filament in place Remove the old filament 4 Insert t...

Page 16: ...op of the flow meter Use a pipe cleaner to clean and free up the bead c Or try this i Disconnect all the power serial and exhaust lines from MLogger Once the box is free from any tether turn the unit...

Page 17: ...low diagram b If all fittings are tight and no leaks are found then the regulator needs to be replaced Switch to the spare MLogger and contact the MSI office 3 Look inside the MLogger on the Red Side...

Page 18: ...is carrier air pressure and the GC flow is at 0 5 SCFH then the Chromatograph Valve Manifold may be malfunctioning Check the 1 8 poly flow fittings for any loose connections or holes in the line If no...

Page 19: ...ence from an electromagnetic field a Check for any fuzz buster HAM radio antenna or satellite phone in operation close by b Cell phones in use next to the MLogger can sometimes affect the filament sig...

Page 20: ...toolbar ii Check the WITS Input bullet is marked iii Check the WITS Interface Be sure the correct EDR company shows in the white field iv Check the WITS port for the correct connection COM 1 in the fr...

Page 21: ...the sever should cache this connection select yes d The black window will then display the various WITS records that MLogger is receiving from the EDR e The TAG column in this screen corresponds to t...

Page 22: ...lashing green lights on both the processor and router then everything may be working fine The small LED for the hard drive light may be broken or has become unplugged from the processor This will not...

Page 23: ...d drive 1 Check how full the Log file and the Chromatograph file are a Look to the Status window in MControl If the LF and the CF are both at 100 the data has wrapped and MLogger is over writing data...

Reviews: