Micro Preamplifier Manual
10
3.3.2
ME/W-SG Test
The following screen shots show the results of a test with a ME/W-SG signal generator. Please read
the data sheet for more information.
Connect the ME/W-SG to the
μ
PA32 via Omnetics connector on the bottom of the signal generator.
The Omnetics connector of the
μ
PA is lettered in black and the Omnetics connector of the ME/W-SG
respectively. Please connect the devices in this way that the letters are on one side, facing to the bottom
of the ME/W-SG as shown on the picture.
Switch the dip switch into the desired position and press the control button to start the device.
The example below shows heart signals (ECG ventricle). The Portable ME32-System used for this
test had a total gain of 1000 and a bandwidth of 1 to 5000 Hz.