Accessories
Default accessories of the Wireless2100-System are the signal generator, the remote control to reactivate
the headstage, and the USB charger for the battery.
ME/W-Signal Generator
The
ME/W-System Signal Generator
can generate a number of pre-defined sine waves and artificial
biological waveforms and feed it into all types of headstages. With those artificial signals it is possible
to test the functionality of the complete system without the need to use an actual animal. Please read
the ME/W-SG data sheet in the "Appendix" for connecting and operating the ME/W-Signal Generator.
Charging Set
The charging set for the battery works via any USB connector. To recharge the battery, please connect
the charging set to an USB port of the computer and plug the battery into the back of the charger with
the provided adaptor cable W2100-C-B-Ch. The regular
charger Wireless-C-HC
is used for batteries with
100, 200 or 300 mAh. The charger
Wireless-C-LC
is used for the 30 mAh battery only.
Important
: Charging the 30 mAh battery with the wrong charger can lead to severe damage and even
fire! Do not recharge any battery without supervision!
The red LED on top indicates that the device is on. The yellow LED indicates that the battery is being
recharged. After approximately one hour the green LED indicates the end of process.
Remote Control
The headstages can be switched off with a wireless software command. To reactivate the headstage,
it is either necessary to disconnect and reconnect the battery, or to turn it back on with the remote
infrared remote control Wireless-F. Just point the remote control at the headstage and press the button.
You need to aim at the receiving LEDs next to the battery connector with a distance of 5-10 cm.