MGS Installation and Operation Manual
6
1000-0085 Rev 0
1.2. Technical Specifications
Specification
Description
Power Supply
12-24 VDC, 12-24 VAC 50/60 Hz, 2 W max.
Power consumption (12V): 60mA (EC), 153mA (SC), 136mA (IR)
Power Monitoring
Green LED
Visual Alarm
Red LED
Audible Alarm
Buzzer, enable/disable
Fault Monitoring
Red LED (on); Green LED (off)
Analogue Outputs
4-20 mA; 0-5 V; 0-10 V; 1-5 V; 2-10 V
Relay (Digital) Outputs
1 relay rated 1 A @ 24 VAC/VDC; Delay: 0, 1, 5, or 10 minutes
IP Rating
IP41 (standard); IP66 (optional)
Temperature Rating
Sensor
Standard Housing
IP66 Housing
IR (all)
-20
°
to 50
°
C
-40
°
to 50
°
C
SC (all)
-20
°
to 50
°
C
-40
°
to 50
°
C
EC (all but NH
3
)
-20
°
to 40
°
C
-20
°
to 40
°
C
EC (0-1K ppm NH
3
)
-20
°
to 40
°
C
-40
°
to 40
°
C
Humidity Rating
0-95% non-condensing
Dimensions/Weights per
Enclosure Type (see Note
below)
Housing
Dimensions
Weight
IP41 (standard)
86 x 142 x 53 mm
180 g
IP66 (optional)
175 x 165 x 82 mm
629 g
w/ Splash Guard
175 x 225 x 82 mm
700 g
w/ Remote Sensor
175 x 155 x 82 mm
790 g
w/ Exd Sensor Head
175 x 155 x 82 mm
1185 g
w/PRV Sensor Head
175 x 155 x 82 mm
830 g
w/ Airflow/Duct
175 x 125 x 82 mm
578 g
Exd (ATEX only)
140 x 180 x 90 mm
2234 g
Standard Compliance
UL 61010-1, CSA C22.2 No. 61010-1, IEC 61010-1, EN 61010-1,
EN 55011, EN 50270, FCC Part 15, Subpart B, WEEE RoHS EuP
NOTE:
The hazardous area
Exd Gas Monitor
products are
designed with individually certified Exd main housing enclosures and
certified Exd remote or attached sensor enclosures. The main
housing enclosure and its PCB assembly are also Exd certified, but
the final
Exd Gas Monitor
assemblies (main enclosure and/or
sensor assembly) are not currently Exd certified, but are pending
additional testing.