13
About the trigger mode
Automatic: Even if no trigger conditions are detected, the scopemeter can still collect waveforms. If there are no
trigger conditions, the scopemeter will be automatically triggered and start to collect data after waiting for a certain
period of time. As it’s not correctly triggered, the waveform displayed by the scopemeter will scroll on the screen
due to failure of synchronization. Once legal trigger signals are detected, the waveform will become stable on the
screen. User may use this mode to monitor low-frequency irregular signal or observe the amplitude of signals, such
as waveform of DC power etc.
Normal: After trigger signals are detected, the scopemeter will start collecting waveform data. If there is no
triggering, the socpemeter will not collect new waveforms. The content of monitor will not be refreshed.
Single : In single-time mode, once trigger conditions are detected, the scopemeter will start to collect waveform
data. After new data is collected, the latest waveform will be automatically saved.
Prompt message of scanning status
Auto
Automatic mode. Waveform can be collected even without trigger conditions.
Ready
Waiting for trigger conditions.
Trig’d
Trigger conditions has been detected.