1)
Turn the scale OFF, Slide the switch on the bottom of the scale from "Norm" to "CAL". Then Place the
scale on a Flat, very stable surface (Be certain the scale is OFF)
2)
Press the
key, a random number will appear on the display
3)
Press the
key again, the display will show 2Ero and then 100.0
4)
Gently place a 100 gram weight on the scale and wait 3 seconds.
5)
Press the
key. The display will show CAL, PASS, then it will return to normal weighing and show
100.0.
6)
calibration is complete. Remove the weight, then slide the switch on the bottom of the scale from
"CAL" to "Norm".
INACCURACY / ERROR___________________________________
The primary reasons for inaccuracy or malfunction are low batteries, incorrect calibration, overload or
operating on an unstable surface. Please keep this in mind and maintain and operate your scale
properly. The scale is a precise instrument and must be handled with the upmost care and caution.
FEATURE_______________________________________________
* Power Up Segment Test
When first turning the unit on, all segments of the display will appear as shown below. This display will
remain for approximately 2 seconds and then reset to 0.
1:8:8:8:8
*
Overload
When an applied load exceeds the capacity. "EEEE" will appear on the display. Remove the excessive
load immediately!
!Remember: You can permanently damage the scale by overloading it!
*
Negative Value
Any tared value or a value left in memory will be displayed as a negative number once all weight is
removed from the unit. Press tare to re-Zero the unit.
* Auto off
An auto shut off feature is provided to conserve battery power. The unit will automatically turn off after
approximately 3 minutes of inactivity.