Durascale precision digital pocket scale manual

INACCURACY / ERROR___________________________________

The primary reasons for inaccuracy or malfunction are low batteries, incorrect calibration, air movement, 
overload or operating on an unstable surface. Please keep this in mind and maintain and operate your scale 
properly. The scale is a precise instrument and must be handled with the utmost care and caution.

Overload

When an applied load exceeds the capacity." ------" will appear on the display. Remove the excessive load 
immediately!

!Remember: You can permanently damage the scale by overloading it!

Negative Value

Any tarred value or a value left in memory will be displayed as a negative number once all weight is removed 
from the unit. Press tare to re-Zero the unit.

* Auto off

An adjustable auto shut off feature is provided to conserve battery power. To program the auto-off time please 
follow these instructions:

1)

 Starting with the scale powered off, press and hold both the ‘ON/OFF’ and ‘PCS’ keys until ‘A.OFF.x’ is 

shown, (x=1,2,or 3,) , this means the auto-off time is 1,2 or 3 minutes). 

2)

 Press ‘PCS’ , the value for ‘x’ will change and flash.

3)

 When you have set the desired value for the auto-off time, press ‘TARE’ to save the setting and then press 

‘ON/OFF’ to exit the auto off time programming mode

* DISPLAY DATA HOLD (LOCK) FUNCTION SET

This advanced scale is equipped with a display lock setting to be used when you wish to keep a displayed 
value on the display. This works well when counting or reloading. The hold system works by not allowing the 
scale to register small weight changes when the hold is switched on.

1) Starting with the scale powered off, press and hold both the ‘ON/OFF’ and ‘UNIT’ keys until ‘Hld.on’ (display 
data hold function is on), or ‘Hld.oF’ (display data hold function is off) appears on the display. When the display 
data hold function is on, and the changing of the weight on the pan is /-2d, the display data will not 
change—it will remain held (locked) for 20 seconds after the displayed weight was stable for 5 seconds;

2) Press ‘UNIT’ to change ‘Hld.on’ to ‘Hld.oF’ or ‘Hld.oF’ to ‘Hld.on’

3) When you have set the desired value for the data lock, press ‘TARE’ to save the setting and then press ‘ON/
OFF’ to exit the lock programming mode

 

KEY PAD FUNCTIONS

UNIT

Unit (mode) selection: Press and release the key once to change weight unit You can select g (grams), oz 

http://myweigh.com/support_manual_durascale.html (4 of 6)7/25/2006 8:20:58 PM

Summary of Contents for durascale 50

Page 1: ...th all of our scales but batteries can run low in storage If your scale simply won t turn on while on Battery power this is often caused by loose battery connections Battery prongs terminals are made of metal They must be making good contact with your batteries in order for the scale to power on You can use a paperclip to slightly bend the battery prongs to make them have a better connection Also ...

Page 2: ...oid rough treatment as this may permanently damage the internal sensor and void your warranty Avoid shaking dropping or otherwise shocking the scale This is a precision instrument and MUST BE HANDLED WITH EXTREME CARE IMPORTANT ADVICE Place the item to be weighed on the platform after the stable weight is displayed remove the item immediately This will prolong the longevity and accuracy of this we...

Page 3: ...nstructions below are for each model so if you re not sure try both sets of instructions to see which one will work for your scale Older D 100 Easy Calibration Start with the scale powered off and nothing on the weighing platform 1 press and hold both the ON OFF and TARE keys until CAL 0 appears on the display 2 Press TARE the 0 in CAL 0 will flash 3 When CAL 50 shows on the display Place a 50g ca...

Page 4: ...ge and flash 3 When you have set the desired value for the auto off time press TARE to save the setting and then press ON OFF to exit the auto off time programming mode DISPLAY DATA HOLD LOCK FUNCTION SET This advanced scale is equipped with a display lock setting to be used when you wish to keep a displayed value on the display This works well when counting or reloading The hold system works by n...

Page 5: ...eight value of a container weight is permanently removed for the remainder of the procedure Just turn the scale ON place the tray or container on weighing platform and press the ZERO key The scale will show 0 00 or 0 0 on the display and you can add items to the container tray NOTE When all weight are removed from the weighing tray the tared value of a container will be displayed as a negative num...

Page 6: ...y be solved by following method 1 Prepare a small screwdriver a fine point tweezer 2 Lightly move away the QC paster on the back of the scale using the tweezer now you will see an adjustable resistor from the hole on the scale back shell 3 When the scale is off press and hold down ON OFF key then press down UNIT key until Hld oN or Hld oF is displayed 4 Press Pcs key the scale will display inner c...

Reviews: