background image

TARE is also used to remove the weight of an item from the display - it also sets the 
scale’s display to zero but the original weight is stored in the memory of the scale.  

You can then press NW/GW to show the weight that you tared off.  For example, if you 
turn the scale on,  then place 100 grams of sugar in the scale’s bowl, the scale will show 
100g.  Now press the TARE key, the scale will show 0,  now press NW/GW and the scale 
will show you 100.  You can press NW/GW again to go back to the weighing process.  
You can also repeat this for several weights. For example you can place an Apple in the 
bowl, press the TARE key, the scale will show 0,  place another apple in the bowl,  press 
the TARE key, the scale will show 0.  Now press the NW/GW button and the scale will 
show you the combined weight of both apples.  This process can be repeated for up to 50 
items and is great for recipes or diets that demand exacting measurements.  For additional 
parts/samples counting please see page 4.    

  It’s  best  to  experiment  with  your  scale  and  become  familiar  with  it’s  features.   

The i5000 is a technologically advanced scale.

 

 
                           

 

 

CALIBRATION 

Recalibration is not necessary. 

This i5000 is a professional scale. Unlike most consumer kitchen scales (which are not 
recalibratable) this scale uses a professional load cell with digital auto calibration. It is 
very unlikely that you will ever have to recalibrate this scale. Some professional users 
may want to recalibrate their scale to keep and maintain exact accuracy.  This is because 
the weight of a mass in one part of the world is not necessarily the same in another part. 
Also, with time and use, small mechanical deviations can occur.   In order to calibrate 
this scale you will need a 5000g (5kg) weight or combination of weights. 

How to calibrate 

1. 

Before entering calibration mode, turn the scale on and let it warm up for about 

one minute.  Then turn it off and proceed to the next step below.

 

2. 

Enter calibration mode 

The Scale should be OFF

 (be sure it’s off).  Press and hold the 

[ZERO]

 key. While 

keeping the 

[ZERO]

 key depressed,  Press 

[ON/OFF].  Then release both keys.  

 The 

display will show a random series of numbers. 

3. 

Calibration by the weight 

Wait for the stable indicator to be displayed then Press 

[ZERO]

. The display will show 

"2Ero"

 and then the display will show 5000. 

4.

 Place a 5000g weight on the tray and then press 

[ZERO]

, the display will show 

"CAL", then after 2 seconds, the display will show “PASS”. Calibration in complete. 
Turn the scale OFF, then turn it back ON and test some weight readings. If they are 
inaccurate, perform calibration again. 
 

FEATURES 

 

Auto shut off (optional).           

?

  

Auto calibration 

 

Auto zero tracking                

?  

Auto backlight  

 

Low batter indication            

  

?  

g/oz/dwt conversion 

Summary of Contents for i2500

Page 1: ...My Weigh i2500 i5000 i5000H User Manual ...

Page 2: ...5000 Just like in an Airplane do not use your cell phone near the scale when it is in use While there is No risk of permanent damage to the scale interference can cause an incorrect calibration or temporarily incorrect weight readings CAUTION Avoid placing the scale in direct sunlight this may cause discoloration or malfunction Replace all batteries at the same time Do not replace only a portion o...

Page 3: ...table this scale uses a professional load cell with digital auto calibration It is very unlikely that you will ever have to recalibrate this scale Some professional users may want to recalibrate their scale to keep and maintain exact accuracy This is because the weight of a mass in one part of the world is not necessarily the same in another part Also with time and use small mechanical deviations ...

Page 4: ...ainer bowl The weighing container is place on the scale Display ON OFF key Turns the scale Power on off ZERO key Sets display to zero TARE key Subtracts weight of a container MODE key Changes weighing units N W G W key Switch N W G W or set Pcs key BOTTOM Battery compartment cover Batteries Batteries are not included Battery compartment DISPLAY ZERO indicator Indicates when the scale zero is corre...

Page 5: ...could be 110 120 220 230 or 240Volts 50 60Hz depending on the area where used so please verify that the adaptor is correct 2 Setting up the scale CAUTION Avoid installing the scale in direct sunlight that may cause discoloration or malfunctions Place your scale on a firmweighing table so that the scale is level The scale will not perform accurately when it is not level Place the scale on the firm ...

Page 6: ...r the Sample Size should be either 10 20 50 or 100 pieces The weight of these samples will show on the display 4 Press Mode and select the counting Mode e g count The indicator on the right side of the display should point to COUNT 5 Select the sample size the same as you chose above by pressing the NW GW key the sample size is the same as in step two 10 20 50or100 6 Press the Mode key The scale w...