8
The MyGlucoHealth Wireless Meter and strip
system is one of the most accurate personal
diabetic care products on the market today.
When tested against calibrated laboratory in-
struments, the MyGlucoHealth Meter meets
higher performance standards when com-
pared to many off-the-shelf glucometers.
In glucose concentrations of less than 4.2
mmol/L, MyGlucoHealth tested within plus
or minus 6.5% of laboratory calibration, 89%
of the time and within 7.5%, 100% of the
time. When testing concentrations of greater
than 4.2 mmol/L, MyGlucoHealth was within
5%, 65% of the time, within 10%, 92% of the
time and within 15%, 98% of the time.
The MyGlucoHealth Wireless meter is plasma calibrated. This makes
results directly comparable to HbA1c test results from a laboratory.
Plasma results are approximately 12% higher than whole blood re-
sults. Keep this in mind when comparing MyGlucoHealth test read-
ings to those of whole blood calibrated meters.
A few high-end or professional quality blood glucose meters,
including the MyGlucoHealth Wireless meter, are already
calibrated or set to give a “plasma” test result, making this
conversion unnecessary. For an accurate comparison of your blood
glucose meter to a laboratory result, you must first know how the me-
ter is calibrated.
METER
ACCURACY
note:
When comparing readings between an older meter and MyGlucoHealth
you may see differences due to the higher accuracy standards and plasma
calibration of MyGlucoHealth.