63
Performance characteristics
Slope
0.96
y-intercept
5.99 mg/dL
Correlation coefficient (r)
0.9955
Number of tests
600
Range tested
33.8 –508 mg/dL
Within ± 5 mg/dL
112/168 (66.7 %)
Within ± 10 mg/dL
157/168 (93.5 %)
Within ± 15 mg/dL
167/168 (99.4 %)
Within ± 5 %
255/432 (59.0 %)
Within ± 10 %
404/432 (93.5 %)
Within ± 15 %
431/432 (99.8 %)
Within ± 15 mg/dL and within ± 15 %
598/600 (99.7 %)
The performance of mylife Aveo
™
blood glucose monitoring
system has been evaluated in laboratory and in clinical tests.
Accuracy
The accuracy of the mylife Aveo
™
blood glucose monitoring
system was assessed by comparing blood glucose results
obtained by patients with those obtained using a YSI Model
2300 Glucose Analyzer, a laboratory instrument.
The following results were obtained by diabetic patients at
clinic centers.
System accuracy results for glucose concentration
<
100 mg/dL
System accuracy results for glucose concentration
≥
100 mg/dL
System accuracy results for glucose concentrations between
33.8 and 508 mg/dL