11. EMC Guidance
Warning: It is not safe to operate the device in
the presence of HF surgical equipment or places
where the intensity of EM disturbances is high.
Warning: Use of this equipment adjacent to or
stacked with other equipment should be avoided
because it could result in improper operation. If
such use is necessary, this device and other equi-
pment should be observed to verify that they are
operating normally.
Warning: Use only accessories and cables
provided by the manufacturer. Accessories and
cables not recommended may affect the EM
emissions or immunity.
Guidance and Manufacturer’s Declaration
- Electromagnetic Emissions
Emissions test
Compliance
RF emissions CISPR 11
Group 1
RF emissions CISPR 11
Class A
Harmonic emissions IEC 61000-3-2
Class A
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Complies
Warning: Maintain minimum distance of 12 inches (30 cm) between any part of the working table and portable
RF communication equipment (including peripherals such as antenna cables and external antennas.) Perfor-
mance of the table might degrade if proper distance is not maintained.
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
Immunity Test
IEC 60601-1-2
Test level
Compliance level
Electrostatic discharge
(ESD) IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±kV, ±15 kV air
±8 kV contact
±2 kV, ±4 kV, ±kV, ±15 kV air
Electrical fast transient
/burst IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
±2 kV for power supply lines
±1 kV for input/output lines
Surge IEC61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Voltage dips, short
interruptions and voltage
variations on power supply
input lines IEC 61000-4-11
0% U; 0.5 cycle T at 0º, 45º,90º,
135º, 180º, 225º, 270º, 315º
0% UT; 1 cycle
70% UT; 25/30 cycle
0% U; 250/300 cycle
0% U; 0.5 cycle T at 0º, 45º,90º,
135º, 180º, 225º, 270º, 315º
0% UT; 1 cycle
70% UT; 25/30 cycle
0% U; 250/300 cycle
Power frequency (50/60 Hz)
Magnetic field IEC61000-4-8
30 A/m
50Hz/60Hz
30 A/m
50Hz/60Hz
Conducted RF
IEC61000-4-6
3 V r.m.s.
150 kHz to 80 MHz
6 V RMS in the ISM bands
between 0.15 MHz and 80 MHz
3 V r.m.s.
150 kHz to 80 MHz
6 V RMS in the ISM bands
between 0.15 MHz and 80 MHz
Radiated RF IEC61000-4-3
3 V/m
80 MHz - 2,7 GHz
80% AM at 1 kHz
3 V/m
80 MHz - 2,7 GHz
80% AM at 1 kHz
NOTE: UT is the a.c. mains voltage prior to application of the test level.