Setup
6 – 5
Setting the foot switch
The setting can be changed from pressing the foot switch to start (ON) or release to stop (OFF), to
pressing the foot switch to start (ON) or pressing it again to stop (OFF).
Press the DOWN key to change the setting value.
The Indicator shows the following order
"
"
Ѝ
"
"
Ѝ
"
"
Ѝ
...
Motion operation (depressing the foot switch turns ON
and releasing the foot switch turns OFF) (factory default
setting)
Alternate operation (press the foot switch to start up,
press it again to stop off)
6 – 6
Setting the handpiece heat protection function
The heat protection function can be set to work if an abnormal heat is detected on the tip of the tool.
Press the DOWN key to change the setting value.
The Indicator shows the following order
"
"
Ѝ
"
"
Ѝ
"
"
Ѝ
...
unprotected
Protected (default setting)
6 – 7
Initializing Settings
You can restore the factory default settings.
While pressing the UP key and the DOWN key (1) at the same time, turn on the power switch (2).
Release the UP key and DOWN key after "beep" and a beep sounds to initialize the user settings. The
output display is set to MIN and the unit starts in normal mode.
②
①
Factory Default Settings
Forced stop time when not
in use
3 minutes.
Notification volume
Notification Sound Loud
Amplitude display and
WATT display
amplitude indication
Foot switch
Motion operation
(depressing the foot switch
turns ON and releasing it
turns OFF)
handpiece heat protection
function
protected species