background image

Instruction 

Manual

DigitalBio is brand name of NanoEnTek.

Summary of Contents for ADAM-rWBC

Page 1: ...Instruction Manual DigitalBio is brand name of NanoEnTek...

Page 2: ...information in this manual is described as accurately as possible Firmware and software changes and updates may change without prior consent or notification Copyright 2013 by NanoEnTek Inc All rights...

Page 3: ...l requirements Power on and initial display Icon functions General operation Preparing the test sample Sample preparation and testing Operating the ADAM rWBC Start input Result analysis Maintenance an...

Page 4: ...r transfusion Targeting the leukocyte the ADAM rWBC uses fluorescent dye to per form cell dyeing and automated cell counting Automated cell counting eliminates user bias or subjective interpretation t...

Page 5: ...which is based on fluo rescent microscopy techniques for counting cells The ADAM rWBC uti lizes sensitive fluorescence dye staining LED optics and CCD detec tion to make provide accurate and reliable...

Page 6: ...are stained while plate lets and red blood cells without nuclei are not stained The ADAM rWBC Kit consists of r Solution and the disposable 1 test r Slide For absolute counting of residual leukocytes...

Page 7: ...ncurred during transit must be filed with the carrier C A U T I O N Neglecting to remove any or all shipping brackets or foam inserts prior to operation may result in damage to the equipment The shipp...

Page 8: ...l procedures of automatic counting Protects the alignment of stage from external shock when the ADAM rWBC is being moved Lock ADAM rWBC before turning it off or moving it Slide is inserted and ejected...

Page 9: ...ad port RS232C port Video port Fans Power switch Fans Power switch Power plug USB port RS232C port Keypad port ADAM rWBC s cooling ADAM rWBC ON OFF power Connects the ADAM rWBC power cord to wall outl...

Page 10: ...or other corrosive substances Area with very little dust or other airborne particles Area with a 10 cm minimum space around the instrument for proper air flow Do not place anything on top of ADAM rWB...

Page 11: ...r Slide is inserted contact your local distributor or ivdst nanoentek com Error Messages during booting System State It appears when booting not working properly Turn off main power and re start devic...

Page 12: ...lation Power on and initial display Icon functions Features Shows when cell counting is started After pressing the Start button Shows the slide holder is ejected After pressing the Eject button Shows...

Page 13: ...48 hours after leukocyte reduction The blood sample is mixed and an aliquot removed and diluted with r Solution for staining After staining a sample is loaded into the r Slide and left to settle Afte...

Page 14: ...0 L sample 400 L reagent 5 Load 100 L mixed sample reagent onto a r Slide 6 Wait 4 7 minutes for sample settling 7 Insert r Slide into ADAM instrument and press to start 8 After approximately 3 minute...

Page 15: ...ror Load 100 L of sample into the r Slide and ensure correct filling The instrument will not detect in low or high sample volumes CAUTION Avoid bubbles as they may negatively affect results Operating...

Page 16: ...ce up in the direction of the ar row on the slide The instrument will not detect if slides are inserted incorrectly See pictures below for proper insertion Correctly inserted Upside down inserted Not...

Page 17: ...The number of total leukocytes will be automatically calculated by the ADAM rWBC software Start input Result analysis General operation Number of total leucocytes 1 71 L Error message during result an...

Page 18: ...enance be cause ADAM rWBC has no consumable materials to replace Frequent or before every testing please clean the exposed outer surface of ADAM rWBC using a soft cloth and isopropyl alcohol or deioni...

Page 19: ...ult error message Solution Check power switch on back of unit Check power source Replace or repurchase Refer to the 11 page Refer to the 11 page Refer to the 12 page Refer to the 17 page Cause Power s...

Page 20: ...recommend that you save the original packing material in which the instrument was shipped This warranty should be limited to the replacement of defective products For any inquiry or request for repair...

Page 21: ...9 Kg Size W L H 220 375 250 mm Degree of protection IPX0 r Slide Measuring range 1 100 cells L Measuring time 2 5 3 min test Loading sample vol 100 L r Slide Measuring vol about 57 L r Slide Solution...

Page 22: ...ntek com This Instruction Manual can be downloaded from the NanoEnTek website www nanoentek com Product list Cat No ADAM rWBC AD1K 050 Contents Main device 50 pcs r Slide RDS 50 r Solution 25mL RDR 50...

Page 23: ...metallic into the air vents of the instrument as this could result in electrical shock personal injury and equipment damage Always set the main switch on the power supply unit to OFF before connectin...

Page 24: ...objet dans les orifices d aeration de l instrument car cela pourrait entrainer un choc electrique des blessures chez les utilisateurs et des dommages d equipement Branchez le cordon d alimentation fe...

Page 25: ...onic products should be disposed of separately from the municipal waste stream via designated collection facilities appointed by the government or the local authorities 3 The correct disposal of your...

Page 26: ...to an authorized service person Do not remove cover or disassemble case There are no adjustable components inside the instrument If a malfunction is found refer to an authorized service person Do not...

Page 27: ...tion or technical assistance please call or email NanoEnTek Inc 851 14 Seohae ro Paltan myeon Hwaseong si Gyeonggi do 18531 Korea Tel 82 2 6220 7942 Fax 82 2 6220 7999 NanoEnTek America Inc 240 Bear H...

Page 28: ...6220 7942 Fax 82 2 6220 7999 NanoEnTek America Inc 240 Bear Hill Road Suite 101 Waltham MA 02451 USA Tel 1 781 472 2558 Fax 1 781 790 5649 EC Representative MT Promedt Consulting GmbH Altenhofstrasse...