Instruction Manual

Automated Somatic Cell Counter

Copyright NanoEnTek, Inc. All rights reserved.

ADAM

SCC

2

Summary of Contents for ADAM-SCC2

Page 1: ...Instruction Manual Automated Somatic Cell Counter Copyright NanoEnTek Inc All rights reserved ADAM SCC2...

Page 2: ......

Page 3: ...14 Seohae ro Paltan myeon Hwaseong si Gyeonggi do 18531 Korea Tel 82 2 6220 7940 Fax 82 2 6220 7999 NanoEnTek America Inc 240 Bear Hill Road Suite 101 Waltham MA 02451 USA Tel 1 781 472 2558 Fax 1 781...

Page 4: ......

Page 5: ...Safety precautions Safety symbols Warnings Technical Support General Operation Introduction Sample preparation Counting cell Measure Run sample Result analysis Result analysis Error code Data Data lis...

Page 6: ...unts the number of somatic cells to determine the quality of raw milk with unique image cytometer technology ADAM SCC2 can be used as an ideal equipment in the research field dairy farms milk manufact...

Page 7: ...curacy precision and reproducibility of this method ADAM SCC2 is fully compatible with the DMSCC It is com posed of the disposable plastic microchips and staining solutions a fluo rescence microscopic...

Page 8: ...e and 100 L stain solution in the 1 5 mL tube Once the experiment is complete the results can be printed through the optional thermal print Printed number indicates cell concen tration x 1000 mL in ea...

Page 9: ...re included with the product Any damage claims must be filed with the carrier The warranty does not cover in transit damage Upon receipt store SOMA Chip at room temperature SCC Solution should be stor...

Page 10: ...erted and ejected Power Power on off LCD Display processes and results START Performs all procedures of automatic counting LOCK Protects the alignment of stage from external shock when ADAM SCC2 is be...

Page 11: ...ear view showing various parts of the ADAM SCC2 Port Description USB Port Port for software update and save the data Power Plug Connects ADAM SCC2 power cord to wall outlet Product Description USB Por...

Page 12: ...7 Ensure minimal contact with dust or other airborne particles 8 Allow a 10 cm 4 inches minimum space around the device for proper airflow 9 Do not place any objects on the device 1 Check the connecti...

Page 13: ...f X axis sensor 0x00007000 Failure of Y axis sensor 0x00008000 Failure of Z axis sensor 0x06000000 Failure of Locking module sensor Getting Started It appears when booting is not working properly Turn...

Page 14: ...Somachip SOMA Chip Set the SOMA Chip according to you are using Cell size Set the minimum and maximum size of cell Dilution factor When diluting sample set the Dilution factor Factor values for the SC...

Page 15: ...hip for automated somatic cell count using the ADAM SCC2 Please check the procedure of sample preparation and testing below For more detailed information please refer to the next page 2 Load the mixed...

Page 16: ...f raw milk sample and 100 L of SCC Solution in the tube 1 1 ratio 2 Mix the raw milk sample and SCC Solution by turning the tube upside down 3 5 times 3 Load the cell sample onto the chip Then wait 1...

Page 17: ...ell Sample loading error Be cautious of loading the correct volume of the sample into SOMA Chip The instrument will not detect low or high sample volumes WARNING T1 N1 T2 N2 Correct volume Low volume...

Page 18: ...slide The instrument will not detect if slides are inserted incorrectly See pictures below for proper insertion WARNING PleaseinsertorremovetheSOMAChipwhentheslideholderisfullyejected Whenthetestisfi...

Page 19: ...he slide Click the title section then scan the cow ID with a barcode scanner to au tomatically enter as the title Start counting process by pressing START It may take about 2 minutes longer for auto f...

Page 20: ...ion Measure Result Analysis The result will be displayed after being automatically calculated by ADAM SCC2 software No Title Count T1 120 1 04 x 10E6 N1 125 1 15 x 10E5 T2 130 4 68 x 10E5 N2 143 2 42...

Page 21: ...e Result Analysis Error code Error code Cause E Frames with errors are over 50 of total counting frame Error frame Frame with error is a frame that contains cells whose diameter is larger than 100 m W...

Page 22: ...y the data of the selected date EDIT View and edit the data Multiple data can be edited with the same settings IMAGE Check the cell images of each channel SAVE Save the selected data to USB PDF Excel...

Page 23: ...on Sample Edit the sample name Cell size graph View the cell size graph Cell size setting Set the min max size of the cell Dilution Factor Set the dilution factor of sample Factor values for the SCC S...

Page 24: ...ction Data IMAGE Control buttons Description Original Check the original image Counted Check the counted cell image Frame Select a frame number of the channel Zoom in out Zoom in and out to check the...

Page 25: ...address from the list to send the Excel PDF and Image files of selected data The e mail address that received the files remains in the list Mail Address If you want to receive files by new e mail ple...

Page 26: ...before counting Refer to page 10 for more information Dilution factor Set the dilution concentration of the sample Wifi Set the wifi to use the e mail function Mail Do not change the setting in mail U...

Page 27: ...update file 2 Insert the USB 3 Click the UPDATE button If connection error occurs please contact a laboratory facility manager CAUTION The AdamUpdate foldermustexistintherootpathoftheUSBfolder ADAM SC...

Page 28: ...there is no operation for 3 minutes the lock function will be activated automatically When the device is locked the screen will be changed as shown below If you press the power button 2 3 seconds the...

Page 29: ...s no replacement of consumable materials 3 Please clean the exposed surface of ADAM SCC2 frequently or before testing using a soft cloth and isopropyl alcohol or deionized water Dispose of wipes in an...

Page 30: ...Adjust the number of cells between 0 05 1 15 x 106 cells mL refer to page 4 Check the expired date Try again after vortexing the cells When error message is shown For information on each error message...

Page 31: ...of fittings or spare parts supplied by anyone other than NanoEnTek 5 Damage caused by accident or misuse 6 Damage caused by disaster 7 Corrosion caused by improper solvent or sample For your protectio...

Page 32: ...4 X 4W Green LED CMOS camera Excitation filter Dichroic filter Emission filter 7 Kg 227 276 270 mm IPX0 23 L test SOMA Chip 2X 13 L test SOMA Chip 4X 8 6 L test SOMA Chip 2X 3 4 L test SOMA Chip 4X 2...

Page 33: ...uct Contents Quantity CRS K01 SOMA Chip 2X Kit 50 pcs SOMA Chip 2X 1 25 mL SCC Solution 1 CRS K02 SOMA Chip 4X Kit 100 pcs SOMA Chip 4X 1 25 ml SCC Solution 2 ADB 500 ADAM Calibration Bead 5 mL Calibr...

Page 34: ...objects especially metallic into the air vents of the instru ment as this could result in electrical shock personal injury and equip ment damage Always set the main switch on the power supply unit to...

Page 35: ...ts Instruments bearing the TUV symbol are certified by TUV SUD America Inc to be in conformance with the applicable safety standard for the US and Canada LED USB Connection When this crossed out wheel...

Page 36: ...n Cover Do not remove cover or dissemble case There are no adjustable components inside the instrument If a malfunction is found refer to an authorized service person Manual Do not attempt to service...

Page 37: ...nical resources including manuals FAQs etc Technical support contact information Additional product information and special offers NanoEnTek Inc 851 14 Seohae ro Paltan myeon Hwaseong si Gyeonggi do 1...

Page 38: ......

Page 39: ......

Page 40: ...eonggi do 18531 Korea Tel 82 2 6220 7940 Fax 82 2 6220 7999 NanoEnTek America Inc 240 Bear Hill Road Suite 101 Waltham MA 02451 USA Tel 1 781 472 2558 Fax 1 781 790 5649 Email sales nanoentek com Webs...