JuLI™ Stage STAT Software
JuLI™ Stage - User Manual
97
The initial image view always displays the
Default
Z-plane, which is 0,0 µm.
In order to pick one Z-plane regarded as optimal, select one channel and scroll
through the stack using the arrow buttons.
Checking the circles
Every Cycles
and
All Wells
followed by
Apply
will use
only this plane of the selected channel in further analyses.
Clicking
Auto Apply
will automatically search for the best plane based on the
contrast of the image in the selected channel.
Roll Back
will reset changes back to the 0,0 µm plane.
If no plane was selected the analysis will be executed on the 0,0 µm plane.
Clicking
Z-Stack
will hide the
Z-Stack
window and turn the button dark blue.
☞
Note: The information which plane was selected will not be accessible later on in
results tables or anywhere else.
⑩
Project information
Once a measurement / project is selected in the list, acquisition details such as
objective lens used, total time, number of cycles and interval time are displayed in
this window. Also, channel settings of the measurement (exposure time, brightness
and LED power) are detailed:
Summary of Contents for JuLi Stage