PainShield® MD User Manual
24
Immunity test
IEC 60601-1-2
Test level
Compliance
Level
Electromagnetic Environment –
Guidance
Surge
IEC 61000-4-5
±1 kV
differential
mode
±2 kV
common
mode
±1 kV
differential
mode
±2 kV
common
mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips,
short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in
U
T
) for 0.5
cycle
40% U
T
(60% dip in
U
T
) for 5
cycles
<5% U
T
(>95% dip in
U
T
) for 5 sec
<5% U
T
(>95% dip in
U
T
) for 0.5
cycle
40% U
T
(60% dip in
U
T
) for 5
cycles
<5% U
T
(>95% dip in
U
T
) for 5 sec
Mains power quality should be
that of a typical commercial or
hospital environment. The use
of PainShield® MD during
continued operation will not be
affected by power mains
interruptions, since PainShield®
MD is battery powered. During
system charging, it is
recommended that PainShield®
MD be powered by an
uninterruptible power supply.
U
T
is the A/C mains voltage prior to application of the test level.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
Portable and mobile RF
communications equipment
should be used no closer to any
part of the PainShield, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the
transmitter.
Recommended separation
distance
Conducted RF
IEC 61000-4-6
3 Vrms
150kHz to
80MHz
[V1] V
P
V
d
1
5
.
3
Radiated RF
IEC 61000-4-3
3 Vrms
80MHz to
2.5GHz
[E1] V/m
P
E
d
1
5
.
3
80 MHz to 800
MHz
P
E
d
1
7
800 MHz to 2.5
GHz