16
8 Specifications
Model
BP 4.0
Measuring Range
Systole (SYS) 70 to 260 mmHg
Diastole (DIA) 30 to 180 mmHg
Pulse (PULSE) 40 to 199 (1/min)
Operating Temperature
+ 10°C to + 40°C + 50°F to + 104°F
Operating Humidity
15 to 90% RH
Storage Temperature
-20°C to 60°C -4°F to + 140°F
Storage Humidity
15 to 95% RH
Indicating Range
Pressure : 20 to 280 mmHg
Pulse: 40 to 199 beats per minute (1/min)
Accuracy
Pressure: ± 3 mmHg or 2% of reading
Pulse: ±5% of reading
Measuring Method
Oscillometric method
Default Cuff Size
Upper arm circumference: 25-35cm 9,5”to12,6”
Cuff Tube Lenght
35 7/16” (600mm)
Inflation
Automatic by electric pump
Deflation
Automatic rapid deflation pressure
Pressure Detection
Resistive semi-conductor pressure sensor
Pulse Wave Detection
Resistive semi-conductor pressure sensor
Air Release
Automatic exhaust valve
Memory capacity
10 records in EEPROM when data cannot be transmitted
Display
Liquid crystal display (LCD)
Operating elements
One button inside the main compartment with square symbol
Data transmission
Bluetooth Class II – Bluetooth Class 4