Page 16
4.5 Options / Start screen
On the Start screen, you can use the O
PTIONS
button to make global program settings.
This includes:
Electrode Settings
Select electrode type:
Standard or (Needle (currently not available))
Electrode Check Interval:
The time interval between the automatic impedance
measurements can be changed.
Max. Impedance:
Shows the set warning / limit value of the permissible
impedance of an electrode.
Max. Impedance Difference: Shows the set warning / limit value of the permissible
impedance difference of the electrodes.
See also chapter 5.9, Electrode Check.
Date / Time
To change the time or date, change the corresponding values and confirm with the
A
PPLY
button.
Language
The language used for the device can be set by selecting the corresponding button.
Marker
Up to 20 different marker texts can be defined in order to access markers with the
predefined texts during the measurement.
Configuration
•
User Preferences:
o
If the option "Show query for channel 1" is activated, the position of the
electrodes of channel 1 (left side of the head / right side of the head /
alternative position) is requested before the start of the measurement.
A corresponding marker is set at the beginning of the measurement.
•
Service: Setup and maintenance menu for service technicians.
•
Device Information: Displays device and version information.