Page 9
2.5 IntelliBridge connector (optional)
The integrated IntelliBridge connector enables the transmission of EEG signals and
evaluated parameters to preconfigured Philips-IntelliVue patient monitors of the series
MP40-90 / MX700 / MX800.
Name and driver version of the Philips EC10 module:
865115/101 IntelliBridge EC10 Module with OpenInterface driver.
The OpenInterface driver must be the revision A.2. The accurate name of the driver
is “ED101-A.2“.
2.6 RS232 connector
The RS232 interface enables the transmission of EEG signals and evaluated
parameters e.g. Narcotrend-Index to store or display the values on an external device
(see also chapter 4.5 options -> configurations).
The reception of data from the external device is not provided.
Pin allocation:
Pin
Signal
2
RS232 TxD (Narcotrend-Compact M)
3
RS232 RxD (Narcotrend-Compact M)
5
RS232 Ground
For further information see the document: “NAT_InterfaceSpecification_2017-09-26”.
•
Integrating the Narcotrend-Compact M into a network can lead to
previously unidentified risks for the patient. These risks should be
identified, analyzed, evaluated and controlled by the responsible
organization.
Changes to the network could introduce new risks that require
additional analysis.
Changes to the network include:
- changes in network configuration
- connection of additional items
- disconnection of items
- update of equipment
- upgrade of equipment
For all Interfaces:
•
For evaluation purposes, primarily the data that is displayed by the
Narcotrend-Compact M shall be used.
•
As data communication errors cannot be excluded, data on the external
device have to be checked with regard to plausibility.
•
USB, RS232 and IntelliBridge are intended to be connected to medical
electrical equipment complying with IEC 60601-1 with SELV.
•
The length of the data cable used may not exceed 3 m.