2.4 Dangers from electromagnetic fields
Narda
NBM-520
9
2.4
Dangers from electromagnetic fields
Strong Fields
Very high field strengths can occur in the vicinity of
some radiation sources.
⇒
Do not cross or ignore safety barriers or markings.
⇒
Persons with electronic implants (e.g. heart pacemak-
ers) must avoid danger zones.
Measurement Errors
Metallic labels (stickers) affixed to the yellow sensor
area of the probe can lead to measurement errors, usu-
ally an underestimation of the electromagnetic field
strength.
⇒
Affix labels of any type only to the black probe shaft.
⇒
If the instrument malfunctions, take it out of service and
contact your Narda Service Center. The addresses are
listed at the end of this operating manual and on the In-
ternet at
http://www.narda-sts.com.
Probe is not operating properly
Possibly present high radiation values are not recog-
nized.
⇒
Check probes for proper operation with a signal source
before using this measuring instrument. This is espe-
cially important for thermocouples because the sensors
can be affected by various mechanical and environmen-
tal stressors.
Narda offers portable sources to accomplish this impor-
tant step (see
⇒
Before beginning any RF radiation measurement, always
inform yourself of the frequencies and field strengths that
you could expect to encounter.
Summary of Contents for nbm-520
Page 1: ...NBM 520 Narda Broadband Field Meter Operating Manual...
Page 6: ...Contents iv NBM 520 Narda...
Page 12: ...1 Useful information 6 NBM 520 Narda...
Page 21: ...3 2 Instrument overview Narda NBM 520 15 3 2 Instrument overview 7 1 2 3 4 9 10 6a 6b 6 5 8...
Page 40: ...4 Measuring with the NBM 520 34 NBM 520 Narda...
Page 52: ...6 Instrument maintenance 46 NBM 520 Narda...
Page 56: ...7 Specifications 50 NBM 520 Narda 7 6 CE Declaration of Conformity...
Page 60: ...8 Ordering information 54 NBM 520 Narda...
Page 63: ......