6.4 Checking for proper operation
Narda
NBM-520
45
Appropriate test sources
The following table shows appropriate test sources for a
function test.
Probe
Test Source:
27 MHz
(2244/90.38)
Test-Source:
446 MHz
PMR Pocket
Radio
1)
1) PMR Pocket Radios are commonly available in electronics stores.
Test-Source:
12 GHz
Model 8699
2)
2) Model 8699 only available for North America.
EF0391
9
9
-
EF1891
9
9
9
EF6091
9
9
9
HF3061
9
9
-
HF0191
9
9
-
EA5091
9
3)
3) LF sensor only
9
9
4)
4) RF sensor only
EB5091
9
9
9
EC5091
9
9
9
ED5091
9
9
9
EF5091
-
9
9
EF5092
-
9
9