8-26
Remote control
?ITG
This query command #?IAE * sends back a string expressing the current internal
Tracking Generator to the mode status. The reply reflects its status and can be:
•
ITG=Off
•
ITG=
TGout
•
ITG=TGin
For details on TG Mode please read the relative documentation.
SIA1 a
This setting command immediately switches then 10dB attenuator according to the
parameter (a) which can be either On or OFF.
The reply is IA1=OK which acknowledges the command has been granted or
IPA =SERR if the command has been ignored.
Example: #SIA1ON*
Example: #SIA0OFF*
SIAE a
This setting command immediately switches 20dB EXTRA Attenuator to the value
indicated by the index (a) as follows:
•
0: 20dB Extra Attenuator OFF
•
1: Enabled 20dB Extra Attenuator 9kHz
•
2: Enabled 20dB Extra Attenuator 150kHz
The reply is IAE=OK which acknowledges the command has been granted or
IAE=SERR if the command has been ignored.
Example: #IAE1* (Switches the 20dB Extra Attenuator 9kHz )
SIAT a
This setting command immediately sets Attenuator to the value indicated by (a)
can range 0 to 30 in step of 10. The reply is IAT=OK which acknowledges the
command has been granted or IAT =SERR if the command has been ignored.
Example: #IAT10* (set attenuator to 10dB)
SIDC a
This setting command immediately switches the CAP-METER according to the
parameter (a) which can be either On or OFF.
The reply is ICM=OK which acknowledges the command has been granted or
ICM=SERR if the command has been ignored.
Example: #SICMON* The CAP-METER is ON
Example: #SIROOFF* CAP-METER is OFF
SIHP a
This setting command immediately switches the 150kHz High-Pass filter according
to the parameter (a) which can be either On or OFF.
The reply is IHP=OK which acknowledges the command has been granted or
IHP=SERR if the command has been ignored.
Example: #SIHPON*
Example: #SIPAOFF*
SIPA a
This setting command immediately switches Preamplifier according to the
parameter (a) which can be either On or OFF.
The reply is IPA=OK which acknowledges the command has been granted or
IPA =SERR if the command has been ignored.
Example: #SIPAON*
Example: #S IPA OFF*
Summary of Contents for PMM FR4003
Page 10: ...X Contents This page has been intentionally left blank...
Page 20: ...1 10 General Information This page has been intentionally left blank...
Page 26: ...2 6 Installation Remove the battery charger and all AC plug adapters enclosed into the bag...
Page 32: ...2 12 Installation This page has been left blank intentionally...
Page 82: ...5 12 Utility Calibration and Analog mode This page has been left blank intentionally...
Page 94: ...7 10 Updating Firmware This page has been left blank intentionally...
Page 134: ...10 4 Packaging Instructions Close the carrying case...
Page 138: ...Suggerimenti Commenti Note Suggestions Comments Note...