Remote control
8-27
SIPS a
This setting command immediately selects the preselector to the index (a) as
follows:
•
0:
OFF
•
1:
9kHz-5.67MHz
•
2:
5.67MHz-11.19MHz
•
3:
11.19MHz-16.71MHz
•
4:
16.71MHz-22.23MHz
•
5:
22.23MHz-30MHz
The reply is IPS=OK which acknowledges the command has been granted or IPS
=SERR if the command has been ignored.
Example: # IPS 1* (Switches the Preselector OFF )
SIRO a
This setting command immediately switches the RF-OUT to BNC connector
according to the parameter (
a
) which can be either
On
or
OFF
.
The reply is
IRO=OK
which acknowledges the command has been granted or
IRO =SERR
if the command has been ignored.
Example:
#SIROON*
The output has been switched to BNC Connector
Example:
#SIROOFF*
The output has been switched to internal ADC
SITG a
This setting command immediately switches the internal Tracking Generator to the
mode indicated by the index (
a
) as follows:
•
0
: TG is OFF
•
1
: TG mode OUT
•
2
: TG mode IN
The reply is
ITG=OK
which acknowledges the command has been granted or
ITG=SERR
if the command has been ignored.
Example:
#ITG1*
(Switches internal Tracking Generator to mode OUT)
For details on TG Mode please read the relative documentation.
Summary of Contents for PMM FR4003
Page 10: ...X Contents This page has been intentionally left blank...
Page 20: ...1 10 General Information This page has been intentionally left blank...
Page 26: ...2 6 Installation Remove the battery charger and all AC plug adapters enclosed into the bag...
Page 32: ...2 12 Installation This page has been left blank intentionally...
Page 82: ...5 12 Utility Calibration and Analog mode This page has been left blank intentionally...
Page 94: ...7 10 Updating Firmware This page has been left blank intentionally...
Page 134: ...10 4 Packaging Instructions Close the carrying case...
Page 138: ...Suggerimenti Commenti Note Suggestions Comments Note...