12 Spatial Averaging
146
SRM-3006
Narda
12.2.1
Continuous
In
Continuous Averaging
mode, the SRM-3006 records the values
continuously and averages them. In this way, particular paths can be tracked
through the room with the antenna. Recording of measurement values is
started by pressing the
Start
key and can be halted by pressing the
Pause
key and then restarted at a different point in the room by pressing the
Continue
key. The measurement continues in the background even if the
instrument is switched to “Pause”; this can be seen by the incrementing of
the “No. of Runs” indicator. These results are not, however, used for the
evaluation. This is indicated by the “No. of SAVG” display remaining static.
Intermediate and final results can be saved at any time by pressing the
SAVE
key.
The SRM-3006 averages all the measured values recorded between “Start”
and “Pause” or “Continue” and “Clear”, regardless of when the
measurement was started or resumed. If several parallel paths are to be
traversed with the antenna, for example, you should try to keep the speed of
movement the same.
Figure 30: Example of continuous recording of values at three points in a room
Example application for Continuous Averaging:
FCC RFR Guidelines (USA)
Start
Continue
Continue
Pause
Pause
Pause
Summary of Contents for SRM-3006
Page 1: ...SRM 3006 Selective Radiation Meter Operating Manual ...
Page 12: ...Contents X SRM 3006 Narda ...
Page 15: ...1 Introduction Narda SRM 3006 3 Figure 1 SRM 3006 in on site use ...
Page 24: ...1 Introduction 12 SRM 3006 Narda ...
Page 50: ...3 Connecting Up and Starting to Use the Device 38 SRM 3006 Narda ...
Page 64: ...4 Operation and Basic Settings 52 SRM 3006 Narda ...
Page 128: ...8 Spectrum Analysis Mode 116 SRM 3006 Narda ...
Page 146: ...10 Scope Mode 134 SRM 3006 Narda ...
Page 168: ...12 Spatial Averaging 156 SRM 3006 Narda ...
Page 178: ...13 Measurement Data Memory 166 SRM 3006 Narda ...
Page 182: ...14 Maintenance and Repairs 170 SRM 3006 Narda ...
Page 186: ...15 PC Software 174 SRM 3006 Narda ...
Page 192: ...16 Remote Control 180 SRM 3006 Narda ...
Page 217: ...17 Specifications Narda SRM 3006 205 17 8 Declaration of Conformity ...
Page 218: ...17 Specifications 206 SRM 3006 Narda ...
Page 224: ...18 Annex A 212 SRM 3006 Narda ...
Page 230: ...19 Ordering Information 218 SRM 3006 Narda ...
Page 238: ...Index 226 SRM 3006 Narda ...
Page 239: ......