6 Functions Common to More Than One Operating Mode
88
SRM-3006
Narda
To change the settings:
1.
Press the
Change Settings
softkey and then change the settings.
2.
Press the
Back to Axis
softkey (highlighted in yellow) to return to
measurement mode.
To finish isotropic measurement:
⇒
Press the
Axis
softkey and select
Single Axis
.
6.7
Adjusting the Y axis
Spectrum • Display • Y-Scale...
Level Recorder • Display • Y-Scale...
Scope • Display • Y-Scale...
There are two ways of adjusting the Y axis:
•
Manual adjustment of Y axis scaling and reference point
•
Automatic adjustment of Y axis
Adjusting the Y axis manually
You can change the scaling as well as the reference point of the Y axis.
Changes you make are applied immediately.
To change the Y axis scaling:
1.
Press the
Y-Scale Range
softkey.
2.
Use the rotary control to highlight the desired value.
ª
The change is applied to the display immediately.
3.
Press the
OK
key or the
ESC
key.
To change the Y axis reference point:
1.
Press the
Y-Scale Ref
softkey.
2.
Use the rotary control to highlight the desired value.
ª
The change is applied to the display immediately.
3.
Press the
OK
key or the
ESC
key.
Adjusting the Y axis automatically
⇒
Press the
Y-Scale Auto
softkey.
ª
The Y axis is adjusted automatically.
In this case, the scaling (Y-Scale Range) is set to 100 dB and the reference
point (Y-Scale Ref) is the same as the measurement range (MR+0 dB).
Summary of Contents for SRM-3006
Page 1: ...SRM 3006 Selective Radiation Meter Operating Manual ...
Page 12: ...Contents X SRM 3006 Narda ...
Page 15: ...1 Introduction Narda SRM 3006 3 Figure 1 SRM 3006 in on site use ...
Page 24: ...1 Introduction 12 SRM 3006 Narda ...
Page 50: ...3 Connecting Up and Starting to Use the Device 38 SRM 3006 Narda ...
Page 64: ...4 Operation and Basic Settings 52 SRM 3006 Narda ...
Page 128: ...8 Spectrum Analysis Mode 116 SRM 3006 Narda ...
Page 146: ...10 Scope Mode 134 SRM 3006 Narda ...
Page 168: ...12 Spatial Averaging 156 SRM 3006 Narda ...
Page 178: ...13 Measurement Data Memory 166 SRM 3006 Narda ...
Page 182: ...14 Maintenance and Repairs 170 SRM 3006 Narda ...
Page 186: ...15 PC Software 174 SRM 3006 Narda ...
Page 192: ...16 Remote Control 180 SRM 3006 Narda ...
Page 217: ...17 Specifications Narda SRM 3006 205 17 8 Declaration of Conformity ...
Page 218: ...17 Specifications 206 SRM 3006 Narda ...
Page 224: ...18 Annex A 212 SRM 3006 Narda ...
Page 230: ...19 Ordering Information 218 SRM 3006 Narda ...
Page 238: ...Index 226 SRM 3006 Narda ...
Page 239: ......