9 Level Recorder Mode
122
SRM-3006
Narda
The maximum values are also shown in addition to the actual or mean
values:
To select a detector type:
1.
Press the
Result Type
softkey.
2.
Use the rotary control to select the desired result type and then press the
OK
key to make the setting.
9.4
Selecting an axis mode (Axis)
Level Recorder • Axis
⇒
See
Selecting an axis mode on page 84
.
9.5
Using noise suppression (Noise Thresh.)
Level Recorder • Meas. Range • Noise Thresh. / Noise Suppr.
⇒
See
Noise suppression (Noise Thresh.) on page 90
.
Table 30: Values displayed in
Level Recorder
mode
Type
Explanation
Peak
Displays the instantaneous measured value.
Note:
The peak value display (Peak) is held for 480 ms. The value
displayed therefore corresponds to the highest level
measured within the last 480 ms. This means that the
numerical display is more stable and the graphical display is
smoothed.
Peak Max
Displays the maximum value that has occurred since the
measurement started.
RMS
The measured values are averaged over a defined period of time.
The resulting value is displayed. A bar indicates the progress of the
measurement until the averaging time is reached. When this time is
reached (complete bar), the instrument refreshes the resulting value
using the new measurement results and discards the oldest results
so that the averaging is
progressive
over time. The averaging time is
defined in the CONF menu.
RMS Max
Displays the maximum value of the averaged values that has
occurred since the measurement started.
Summary of Contents for SRM-3006
Page 1: ...SRM 3006 Selective Radiation Meter Operating Manual ...
Page 12: ...Contents X SRM 3006 Narda ...
Page 15: ...1 Introduction Narda SRM 3006 3 Figure 1 SRM 3006 in on site use ...
Page 24: ...1 Introduction 12 SRM 3006 Narda ...
Page 50: ...3 Connecting Up and Starting to Use the Device 38 SRM 3006 Narda ...
Page 64: ...4 Operation and Basic Settings 52 SRM 3006 Narda ...
Page 128: ...8 Spectrum Analysis Mode 116 SRM 3006 Narda ...
Page 146: ...10 Scope Mode 134 SRM 3006 Narda ...
Page 168: ...12 Spatial Averaging 156 SRM 3006 Narda ...
Page 178: ...13 Measurement Data Memory 166 SRM 3006 Narda ...
Page 182: ...14 Maintenance and Repairs 170 SRM 3006 Narda ...
Page 186: ...15 PC Software 174 SRM 3006 Narda ...
Page 192: ...16 Remote Control 180 SRM 3006 Narda ...
Page 217: ...17 Specifications Narda SRM 3006 205 17 8 Declaration of Conformity ...
Page 218: ...17 Specifications 206 SRM 3006 Narda ...
Page 224: ...18 Annex A 212 SRM 3006 Narda ...
Page 230: ...19 Ordering Information 218 SRM 3006 Narda ...
Page 238: ...Index 226 SRM 3006 Narda ...
Page 239: ......