11 UMTS Mode
Narda
SRM-3006
139
11.3.3
Partial results for certain radio cells
(scrambling codes)
It is often necessary to determine the proportion of the overall field strength
within a UMTS frequency channel that is due to certain cells only. You can
use the
Select Menu
softkey function to select the cells of interest by
means of their scrambling codes.
To select individual cells:
1.
Press the
Select Menu
softkey.
2.
Use the rotary control to highlight items and select them with the
softkeys.
3.
Press the
OK
key to confirm the selection.
The selection applies to all display types and all evaluations, i.e. the
numerical overall result (Total) as well as the graphical display versus time
only take the selected scrambling codes into account.
If you select only one scrambling code, you can display the field strength
versus time for this one cell.
11.3.4
Analog measurement result (Analog)
This value directly shows the result for the selected UMTS frequency
channel, determined by analog measurement using a fixed resolution
bandwidth (RBW) of 5 MHz. The extrapolation factor is not applied to this
result.
11.4
Basic settings
This section describes the basic settings. If these settings are identical in
more than one operating mode, a cross-reference will indicate the relevant
section in Chapter 6
Summary of Contents for SRM-3006
Page 1: ...SRM 3006 Selective Radiation Meter Operating Manual ...
Page 12: ...Contents X SRM 3006 Narda ...
Page 15: ...1 Introduction Narda SRM 3006 3 Figure 1 SRM 3006 in on site use ...
Page 24: ...1 Introduction 12 SRM 3006 Narda ...
Page 50: ...3 Connecting Up and Starting to Use the Device 38 SRM 3006 Narda ...
Page 64: ...4 Operation and Basic Settings 52 SRM 3006 Narda ...
Page 128: ...8 Spectrum Analysis Mode 116 SRM 3006 Narda ...
Page 146: ...10 Scope Mode 134 SRM 3006 Narda ...
Page 168: ...12 Spatial Averaging 156 SRM 3006 Narda ...
Page 178: ...13 Measurement Data Memory 166 SRM 3006 Narda ...
Page 182: ...14 Maintenance and Repairs 170 SRM 3006 Narda ...
Page 186: ...15 PC Software 174 SRM 3006 Narda ...
Page 192: ...16 Remote Control 180 SRM 3006 Narda ...
Page 217: ...17 Specifications Narda SRM 3006 205 17 8 Declaration of Conformity ...
Page 218: ...17 Specifications 206 SRM 3006 Narda ...
Page 224: ...18 Annex A 212 SRM 3006 Narda ...
Page 230: ...19 Ordering Information 218 SRM 3006 Narda ...
Page 238: ...Index 226 SRM 3006 Narda ...
Page 239: ......