6 Functions Common to More Than One Operating Mode
74
SRM-3006
Narda
6.3.2
Automatic measurement range search
(MR Search)
Safety Evaluation • Meas. Range • MR Search
This softkey initiates a search for a suitable measurement range. This is
recommended if the field situation is unknown.
⇒
Press the
MR Search
softkey.
ª
The message:
Searching for best measurement range. Please
wait
. is displayed. A measurement over the entire frequency range
covered by the antenna you are using is made in the background.
This takes several seconds. When it is completed, the device sets the
measurement range to the optimum value automatically and reverts
to normal measurement mode.
Note:
Press the
ESC
key if you want to cancel the process.
MR Search Type
The SRM-3006 provides two types of measurement range search which can
be selected using the
MR Search Type
softkey (toggle function):
•
Normal:
This type is suitable for all largely static signals.
•
Conserv.:
This type allows for a higher overload reserve. It is better to set
the measurement range manually if pulsed signals are being measured.
To select the search type:
⇒
Press the
MR Search Type
softkey.
ª
The switch toggles between the two types each time the softkey is
pressed.
6.3.3
Measurement input overload
If the signal level is higher than the range that can be measured, the warning
message
OVERDRIVEN
is shown in the display (in the part above the lower
status line, next to the axis setting).
Summary of Contents for SRM-3006
Page 1: ...SRM 3006 Selective Radiation Meter Operating Manual ...
Page 12: ...Contents X SRM 3006 Narda ...
Page 15: ...1 Introduction Narda SRM 3006 3 Figure 1 SRM 3006 in on site use ...
Page 24: ...1 Introduction 12 SRM 3006 Narda ...
Page 50: ...3 Connecting Up and Starting to Use the Device 38 SRM 3006 Narda ...
Page 64: ...4 Operation and Basic Settings 52 SRM 3006 Narda ...
Page 128: ...8 Spectrum Analysis Mode 116 SRM 3006 Narda ...
Page 146: ...10 Scope Mode 134 SRM 3006 Narda ...
Page 168: ...12 Spatial Averaging 156 SRM 3006 Narda ...
Page 178: ...13 Measurement Data Memory 166 SRM 3006 Narda ...
Page 182: ...14 Maintenance and Repairs 170 SRM 3006 Narda ...
Page 186: ...15 PC Software 174 SRM 3006 Narda ...
Page 192: ...16 Remote Control 180 SRM 3006 Narda ...
Page 217: ...17 Specifications Narda SRM 3006 205 17 8 Declaration of Conformity ...
Page 218: ...17 Specifications 206 SRM 3006 Narda ...
Page 224: ...18 Annex A 212 SRM 3006 Narda ...
Page 230: ...19 Ordering Information 218 SRM 3006 Narda ...
Page 238: ...Index 226 SRM 3006 Narda ...
Page 239: ......