NI 4065 Calibration Procedure
10
ni.com
6.
Call the niDMM Config Measurement VI. The niDMM Config
Measurement VI is polymorphic. To change the selected instance of
this polymorphic VI, right-click the VI and choose
Select Type
followed by
Resolution in Digits
.
7.
Set a writable niDMM property node to set the input resistance of the
NI 4065 to >10 G
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_ConfigureMeasurement
Digits
with the following parameters:
Instrument_Handle
: The instrument
handle from
niDMM_init
Resolution_Digits
:
6.5
Measurement_Function
:
NIDMM_VAL_DC_VOLTS
Range
:
0.1
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM
SetAttributeVi
Real64
with the
following parameters:
Instrument_Handle
:
The instrument handle
from
niDMM_init
Attribute_ID
:
NIDMM_ATTR_INPUT_
RESISTANCE
Attribute_Value
:
NIDMM_VAL_GREATER
_THAN_10_GIGAOHM