NI 4065 Calibration Procedure
60
ni.com
9.
Call the niDMM Cal Adjust Offset VI.
10. Call the niDMM Cal Adjust Offset VI.
11. Call the niDMM Cal Adjust Offset VI.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustOffset
with
the following parameters:
Instrument_Handle
:
Cal Session
Mode
:
NIDMM_VAL_DC_VOLTS
Range
:
1
Input_Resistance
:
NIDMM_VAL_GREATER_THAN_10
_GIGAOHM
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustOffset
with
the following parameters:
Instrument_Handle
:
Cal Session
Mode
:
NIDMM_VAL_DC_VOLTS
Range
:
10
Input_Resistance
:
NIDMM_VAL_GREATER_THAN_10
_GIGAOHM
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustOffset
with
the following parameters:
Instrument_Handle
:
Cal Session
Mode
:
NIDMM_VAL_DC_VOLTS
Range
:
100
Input_Resistance
:
NIDMM_VAL_10_MEGAOHM