National Instruments Corporation
15
NI 4065 Calibration Procedure
14. Repeat step 13 for each of the remaining iterations shown in Table 3.
15. Reset the calibrator.
16. Call the niDMM Config Measurement VI. The niDMM Config
Measurement VI is polymorphic. To change the selected instance of
this polymorphic VI, right-click the VI and choose
Select Type
followed by
Resolution in Digits
.
Table 3.
DC Voltage Offset Settings
Iteration
niDMM Config Measurement Parameters
Range (V
dc
)
Input Resistance
1
1
NIDMM_VAL_GREATER_THAN_10_GIGAOHM
2
1
NIDMM_VAL_10_MEGAOHM
3
10
NIDMM_VAL_GREATER_THAN_10_GIGAOHM
4
10
NIDMM_VAL_10_MEGAOHM
5
100
NIDMM_VAL_10_MEGAOHM
6
300
NIDMM_VAL_10_MEGAOHM
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_ConfigureMeasurement
Digits
with the following parameters:
Instrument_Handle
: The instrument
handle from
niDMM_init
Resolution_Digits
:
6.5
Measurement_Function
:
NIDMM_VAL_DC_VOLTS
Range
:
0.1