NI 4065 Calibration Procedure
34
ni.com
f.
Call the niDMM Read VI.
g.
Verify that this measurement falls between the tolerances listed in
Table 22. Tolerances are provided instead of absolute limits,
because your calibrator will have different discrete resistance
values.
6.
Repeat step 5 for each of the remaining iterations listed in Table 8.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the following
parameters:
Instrument_Handle
: The instrument
handle from
niDMM_init
Reading
: Verify that this measurement
falls between the tolerances listed in
Table 22
Maximum_Time
: –
1
Table 8.
4-Wire Resistance Settings
Iteration
Range (
)
Calibrator
Output
Delay (seconds)
1
100
0
0.0
2
10
0.0
3
100
0.0
4
1 k
0
0.0
5
100
0.0
6
1 k
0.0
7
10 k
0
0.0
8
1 k
0.0
9
10 k
0.1
10
100 k
0
0.0
11
10 k
0.1
12
100 k
0.25
13
1 M
0
0.0
14
100 k
0.25
15
1 M
1.0