NI 4065 Calibration Procedure
38
ni.com
3.
Wait two minutes for the thermal EMF to stabilize.
4.
Refer to Table 10 for the appropriate calibrator output and function
parameter values as you complete the following steps:
a.
Call the niDMM Config Measurement VI. The niDMM Config
Measurement VI is polymorphic. To change the selected instance
of this polymorphic VI, right-click the VI and choose
Select Type
followed by
Resolution in Digits
.
Table 9.
Fluke 5440 Cable Connections
Banana Plug Connector
(NI 4065)
Banana Plug Color
(Fluke 5440 Cable)
Binding Post Label
(Fluke 5720A Calibrator)
HI
Red
OUTPUT HI
LO
Black
OUTPUT LO
(No connection)
Blue
V GUARD
HI
Red
OUTPUT HI SENSE
LO
Black
OUTPUT LO SENSE
(No connection)
Blue
V GUARD
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_ConfigureMeasurement
Digits
with the following parameters:
Instrument_Handle
: The instrument
handle from
niDMM_init
Resolution_Digits
:
6.5
Measurement_Function
:
NIDMM_VAL_2_WIRE_RES
Range
: The Range as shown in Table 10
for the current iteration