National Instruments Corporation
7
NI 4065 Calibration Procedure
4.
Call the niDMM Configure Powerline Frequency VI.
Verification Procedures
You can use the verification procedures described in this section for both
pre-adjustment and post-adjustment verification. The steps of each
verification procedure must be performed in the order listed; however, you
can omit entire sections (for example, the entire
section), if necessary. Refer to
Appendix A: Calibration Options
for more
information.
The parameters
Range
and
Resolution in Digits
used in VI and function
calls throughout this section have floating point values. For example,
if
Range
=
1
, the floating point value is
1.0
. Refer to the
NI Digital
Multimeters Help
for more information about parameter values.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_ConfigurePowerLineFrequency
with the following parameters:
Instrument_Handle
: The instrument handle
from
niDMM_init
PowerLine Frequency
: Set this parameter to
50
or
60
, depending on the powerline
frequency (in hertz) powering your
instruments; select
50
for 400 Hz powerline
frequencies