NI 4065 Calibration Procedure
8
ni.com
Verifying DC Voltage
To verify DC voltage of the NI 4065, complete the following steps:
1.
Reset the calibrator.
2.
Connect the NI 4065 and the Fluke 5700A/5720A calibrator using the
Fluke 5440 cable, as shown in Figure 1 for NI PXI/PCI/PCIe-4065 and
Figure 2 for NI USB-4065. Table 2 lists the cable connections.
Figure 1.
NI PXI/PCI/PCIe-4065 Cable Connections for Voltage and 2-Wire Resistance
1
NI PCI-4065
2
Fluke 5700A/5720A Calibrator
3
Fluke 5440 Cable
CAT II
INPUT
300V
HI
LO
AMPS
3A
SENSE
300V
AUX
I/O
5V
HI
LO
HI
LO
HI
LO
HI
AUX
CURRENT
GUARD
GROUND
1
2
3
HI
HI