NI 4065 Calibration Procedure
Figure 2.
NI USB-4065 Cable Connections for Voltage and 2-Wire Resistance
3.
Wait two minutes for the thermal EMF to stabilize.
4.
Generate 0 V on the calibrator.
Allow the calibrator output to settle
before proceeding
.
5.
Call the niDMM Reset VI.
1
NI USB-4065
2
Fluke 5700A/5720A Calibrator
3
Fluke 5440 Cable
Table 2.
Fluke 5440 Cable Connections
Banana Plug Connector
(NI 4065)
Banana Plug Color
(Fluke 5440 Cable)
Binding Post Label
(Fluke 5700A/5720A Calibrator)
HI
Red
OUTPUT HI
LO
Black
OUTPUT LO
(No connection)
Blue
V GUARD
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_reset
with the following
parameter:
Instrument_Handle
: The instrument
handle from
niDMM_init