Calibration Procedure | © National Instruments | 3
Test Equipment
National Instruments recommends that you use the equipment in Table 1 for calibrating the
NI 449x.
Table 1.
Recommended Equipment
Equipment
Recommended
Model
Parameter
Measured
Minimum Requirements
Calibrator
Fluke 5700A
AI AC
coupled
gain
AI flatness
Frequency Range:
20 Hz to 92.2 kHz
Voltage Range: up to 9 V
pk
AC Accuracy: ±0.05%
*
,
20 Hz to 95 kHz
Function Generator
†
Agilent 33250A
Timebase
frequency
†
Frequency Range: up to
10 kHz
Frequency Accuracy:
±2 ppm
*
Voltage Range: up to 9 V
pk
Voltage Source
NI PXI-6233
AI DC
coupled
gain
Voltage Range: ±10 V
Output isolated from
EARTH ground
‡
Digital Multimeter
NI PXI-4070
AI DC
coupled
gain
Voltage Measurement
Accuracy: ±200 ppm
*
InfiniBand to BNC
Cable
NI 4495/4496/4497/
4498/4499: quantity 2
NI PXIe-4492:
quantity 1
NI SHB4X-8BNC
All
N/A
BNC Shorting Cap
NI 4495/4496/4497/
4498/4499: quantity 16
NI PXIe-4492:
quantity 8
Pomona
Electronics 5085
AI offset
Resistance:
≤
50
Ω