VXIpc-486 Configuration and Installation
Chapter 2
VXIpc-486 Model 500 Series
2-6
© National Instruments Corporation
MODID Signal Termination
W26 configures the MODID signal termination for Slot 0/Non-Slot 0 operation. In the default
configuration for Slot 0 operation (Figure 2-2a), W26 causes the MODID signal to be terminated
with a 16.9 k
Ω
pull-up resistor. When configured for Non-Slot 0 operation (Figure 2-2b), W26
causes the MODID signal to be terminated with an 825
Ω
pull-down resistor.
W26
P2
P1
W25
W24
W34
W33
W36
W35
W32
W20
J16
W26
W26
a. Slot 0*
* Factory Configuration
b. Non-Slot 0
Figure 2-2. MODID Signal Termination Settings
System Controller
W25 configures the System Controller for Slot 0/Non-Slot 0 operation. In the default
configuration for Slot 0 operation (Figure 2-3a), W25 enables the VXIpc-486 Model 500 Series
System Controller functions—SYSCLK and SYSRESET drivers, the Bus Arbiter and Bus
Timer. When configured for Non-Slot 0 operation (Figure 2-3b), W25 disables the System
Controller functions.
W26
P2
P1
W25
W24
W34
W33
W36
W35
W32
W20
J16
W25
a. Slot 0*
* Factory Configuration
b. Non-Slot 0
W25
Figure 2-3. System Controller Settings