National Instruments Corporation
29
NI 5122/5124 Calibration Procedure
Trigger Sensitivity
Complete the following steps using the entries for iterations 1 through 4
in Table 12 to verify trigger sensitivity for input channel 0.
1.
Call
niScope_ConfigureChanCharacteristics
(niScope
Configure Chan Characteristics VI) with the following parameters:
•
channelList
:
"0"
•
inputImpedance
:
NISCOPE_VAL_50_OHM
•
maxInputFrequency
:
100,000,000
2.
Call
niScope_ConfigureVertical
(niScope Configure Vertical
VI) with the following parameters:
•
channelList
:
"0"
•
range
:
0.2
•
offset
:
0.0
•
coupling
:
NISCOPE_VAL_DC
•
probeAttenuation
:
1.0
•
enabled
:
NISCOPE_VAL_TRUE
3.
Call
niScope_ConfigureHorizontalTiming
(niScope Configure
Horizontal Timing VI) with the following parameters:
•
minSampleRate
:
100,000,000
•
minNumPts
:
1,000
•
refPosition
:
50.0
•
numRecords
:
50
•
enforceRealtime
:
NISCOPE_VAL_TRUE
4.
Call
niScope_ConfigureTriggerEdge
with the following
parameters:
•
triggerSource
: The Trigger source value from Table 12
•
level
: The Level value from Table 12
•
slope
: The Slope value from Table 12
•
triggerCoupling
:
NISCOPE_VAL_DC
•
holdoff
: 0
•
delay
: 0
Note
The trigger level is set to center the trigger hysteresis window at 0.0 V.