NI 5122/5124 Calibration Procedure
8
ni.com
External Calibration Options
External calibration involves both adjustment and verification. Adjustment
is the process of measuring and compensating for device performance to
improve the measurement accuracy. Performing an adjustment updates the
calibration date, effectively resetting the calibration interval. The device is
guaranteed to meet or exceed its published specifications for the duration
of the calibration interval. Verification is the process of testing the device
to ensure that the measurement accuracy is within certain specifications.
Verification can be used to ensure that the adjustment process was
successful, or to determine if the adjustment process needs to be performed
at all. During verification, you must compare the measurement error to
limits given in Tables 3 through 12.
This document provides two sets of test limits for most verification
stages—the
calibration test limits
and the
published specifications
. The
calibration test limits are more restrictive than the published specifications.
If all of the measurement errors determined during verification fall within
the calibration test limits, the device is guaranteed to meet or exceed its
published specifications for a full calibration interval (two years). For this
reason, you must verify against the calibration test limits when performing
verification after adjustment. If all of the measurement errors determined
during verification fall within the published specifications, but not within
the calibration test limits, the device is meeting its published specifications.
However, the device will not necessarily remain within these specifications
for an additional two years. The device will meet published specifications
for the remainder of the current calibration interval. In this case, you can
perform an adjustment if you want to further improve the measurement
accuracy or reset the calibration interval. If some measurement errors
determined during verification do not fall within the published
specifications, you must perform an adjustment to restore the device
operation to its published specifications.
The
section describes the recommended calibration
procedure. The
procedures that allow you to skip adjustment if the device already meets its
calibration test limits or published specifications.